- Measurements include conductivity, resistivity, salinity, total dissolved solids and temperature.
- With YSI’s exclusive accurate Resistance Ratio TechnologyTM (RRT), the 3200 continually compares the resistance of the sample to a calibrated resistor in the circuit.
- Ensure the most accurate measurements with NIST-traceable conductivity cells with built-in temperature sensors and exceed USP24 specifications.
- To ensure even greater accuracy, utilize the YSI conductivity calibration solutions.
Features:
- Highly accurate Resistance Ratio Technology
- Multipoint calibration
- High and low alarms
- Linear and nonlinear temperature compensation options
- Data storage and logging
YSI 3200 Instrument General Specifications
Alarm | Yes | |
Calibration | Multipoint, up to 5 points |
|
Cell Connector | 7-pin Mini Din | |
Cell constant, cm-1 | 0.001 – 100 cm-1 | |
Cell Information Storage |
6 Configurations | |
Certifications | UL, CSA, CE | |
Clock | Yes | |
Data Storage | 100 points | |
Dimensions | Size | 9 by 9.5 by 4.4 in (22.9 by 24.1 by 11.2 cm) |
Weight | 2.6 lbs (1.1 kg) |
|
Display | Graphic LCD | |
Operating Environment | 95% relative humidity, non-condensing | |
Outputs | RS-232 | |
Platinizing Capability | Included | |
Power | AC, 115V, 220V | |
Temperature Compensation |
Method | Linear, nonlinear |
Reference temperature °C | 0 – 100 | |
Temperature. Coefficient | 0 – 10%/°C, nonlinear |
|
Water Resistance |
The Model 5000 and 5100 are designed exclusively for indoor use and are NOT waterproof. | |
Warranty |
2-year instrument |
YSI 3200 Specifications (Instrument Only)
Conductance | |||
Range | Accuracy | Resolution | Frequency |
0 – 0.9999 µS | ± 0.30% full scale | 0.0001 µS | 40 Hz |
0.950 – 9.999 µS | ± 0.20% full scale | 0.001 µS | 80 Hz |
9.50 µS – 99.99 µS |
± 0.10% full scale | 0.01 µS | 290 Hz |
95.0 – 999.9 µS | ± 0.10% full scale | 0.1 µS | 1010 Hz |
950 – 9999 µS | ± 0.10% full scale | 1 µS | 1010 Hz |
9.50 – 99.99 mS | ± 0.10% full scale | 0.01 mS | 1010 Hz |
95.0 – 999.9 mS | ± 0.30% full scale | 0.1 mS | 1460 Hz |
0.95 – 3.00 S | ± 1.0% full scale | 0.01 S | 1460 Hz |
Resistance |
|||
Range | Accuracy | Resolution | Frequency |
0 – 9.999 Ω | ± 0.2% full scale | 0.001Ω | 1460 Hz |
0 – 99.99 Ω | ± 0.1% full scale |
0.01 Ω | 1460 Hz |
0 – 999.9 Ω | ± 0.1% full scale |
0.1 Ω | 1010 Hz |
0 – 9.999 kΩ | ± 0.1% full scale |
0.001 kΩ | 1010 Hz |
0 – 99.99 kΩ | ± 0.1% full scale |
0.01 kΩ | 1010 Hz |
100.0 – 999.9 kΩ | ± 0.2% full scale | 0.1 kΩ | 290 Hz |
1.00 – 9.99 MΩ | ± 0.5% full scale |
0.01 MΩ | 80 Hz |
10.0 – 29.9 MΩ | ± 1% full scale | 0.1 MΩ |
40 Hz |
Range | Accuracy | Resolution | |
Salinity | 0-80 ppt (NaCl) | ±.1 ppt | 0.1 ppt |
TDS | 0-19,999 mg/L | 0.50% | 1 mg/L |
Temperature |
-5 – 100°C | ±0.1°C | 0.01°C |
Conductivity Cell General Measurement Ranges
Conductivity Cell General Specifications
3200 Series Cells with built-in temperature sensors |
||||||||||
Model | cgs Cell Type |
S.I. Cell Constant |
Cell Constant | Material | Overall Length (mm) | Max O.D. (mm) | Chamber I.D. (mm) | Chamber Depth (mm) | Volume | |
A | 3252 | dip | 1.0/cm | 100/m | ABS plastic | 146 |
13 |
10 |
20 |
|
B | 3253 | dip, micro |
1.0/cm | 100/m | Pyrex 7740 | 178 |
13 |
10 |
51 |
|
C |
3254 | fill | 1.0/cm | 100/m | Pyrex 7740 | 135 |
19 |
11 |
83 |
5 mL |
D | 3255 | flow | 0.1/cm | 10/m | Pyrex 7740 | 146 |
25 |
21 |
76 |
30 mL |
E | 3256 | dip | 0.1/cm | 10/m | Pyrex 7740 | 159 |
25 |
21 |
52 |
|
|
||||||||||
Model | cgs Cell Type | S.I. Cell Constant | Cell Constant | Material | Overall Length (mm) |
Max O.D. (mm) |
Chamber I.D. (mm) | Chamber Depth (mm) | Volume | |
F |
3401 |
dip | 1.0/cm | 100/m | Pyrex 7740 | 191 |
25 | 21 | 76 | |
G |
3402 |
dip | 0.1/cm | 10/m | Pyrex 7740 | 159 |
25 | 21 | 52 | |
H |
3403 |
dip | 1.0/cm | 100/m | Pyrex 7740 | 178 | 13 | 10 | 51 | |
I |
3417 |
dip | 1.0/cm | 100/m | ABS plastic | 146 | 13 | 10 | 20 | |
I | 3418 | dip | 0.1/cm | 10/m | ABS plastic | 159 | 13 | 10 | 30 | |
J |
3440 |
dip | 10/cm | 1000/m | Pyrex 7740 | 203 | 13 | 2 | 86 | |
*Requires a YSI 3232 Cell Adaptor for use with YSI 3100 and 3200 Conductivity Instruments. For automatic temperature compensation, use a YSI 3220 on the 3200 |