Specifications (PDF)
525A-P Series Pressure Module Specifications | ||||
Type | Fluke Model | Range/Resolution | Range/Resolution | (180 day) |
Differential | 525A-P02 | 1psi / 0.00001 | 6900 Pa / 0.01 | 0.01% FS |
Gage | 525A-P05 | 30 psi / 0.0001 | 207 kPa / 0.001 | 0.01% FS |
Gage | 525A-P06 | 100 psi / 0.001 | 690 kPa / 0.001 | 0.01% FS |
Gage | 525A-P07 | 500 psi / 0.001 | 3400 kPa / 0.01 | 0.01% FS |
Gage | 525A-P08 | 1000 psi / 0.01 | 6900 kPa / 0.01 | 0.01% FS |
Gage | 525A-P29 | 3000 psi / 0.01 | 20.7 M Pa / 0.0001 | 0.01% FS |
Absolute | 525A-PA6 | 100 psia / 0.001 | 690 KPa / 0.0001 | 0.01% FS |
Absolute | 525A-PA7 | 500 psia / 0.001 | 3400 kPa / 0.01 | 0.01% FS |
Absolute | 525A-PA8 | 1000 psia / 0.01 | 6900 kPa / 0.01 | 0.01% FS |
Vacuum | 525A-PV4 | – 15 psi / 0.0001 | – 103 kPa / 0.001 | 0.01% FS |
Absolute | 525A-PA4 | 15 psia / 0.0001 | 103 kPa / 0.001 | 0.01% FS |
Media – dry, non-corrosive gases | |
Total uncertainty (180 day): 0.01% FS (15°C to 45°C) : 1 year 0.025 % FS (15°C to 45°C) | |
Overpressure: 150 % or greater | |
Compensated temperature range: 15°C to 45°C | |
Confidence interval: 95% | |
General Specifications, 525A | |
Warm-up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
Settling Time | Less than 5 seconds for all functions and ranges except as noted |
Interfaces | Standard, RS-232; Optional, IEEE-488 (GPIB) |
Temperature Performance | Operating: 0 °C to 50 °C; Calibration (tcal): 15 °C to 35 °C; Storage: -20 °C to 70 °C |
Electromagnetic Compatibility | Meets CE requirements, including susceptibility |
Temperature Coefficient | Temperature Coefficient for temperatures outside tcal +5 °C is 10% of the 90-day specification |
(or 1-year, as applicable) per °C | |
Relative Humidity | Operating: <80% to 30 °C, <70% to 40 °C,<40% to 50 °C; Storage: <95%, noncondensing |
Altitude | Operating: 3,050m (10,000 ft) maximum; Non-operating: 12,200m (40,000 ft) maximum |
Safety | Designed to comply with IEC 61010 Second Edition; ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92 |
Has CSA marking or equivalent | |
Analog Low Isolation | 20V |
EMC | Designed to comply with IEC 61326-1/1997 (EMC) |
Line Power | Line Voltage (selectable): 100/120V or 220/240V; Line Frequency: 47 to 63 Hz |
Line Voltage Variation: +10% about line voltage setting | |
Power Consumption | < 10 VA |
Dimensions | Height, 13.2 cm (5.25 inches), plus 1.5 cm (0.6 inches) for feet on bottom of unit; Width, 32.6 cm (12.85 inches), |
3/4 standard rack width; Depth, 33.3 cm (13.1 inches) overall; 23.9 cm (9.4 inches) chassis only | |
Weight (without options) | 9 Lbs (4 Kg) |
Be the first to review “METROLOGY WELL, -30C, W/9171-INSF, 115V”