Accessory | Description |
---|---|
9301 |
Carrying Case, fits Tweener and 12 in probe |
2506 |
IEEE Option |
2505 |
Spare Connector |
2508 |
Serial Cable Kit |
9320A |
External Battery Pack (powers 115 V ac or 12 V dc products only) |
SC1-THRO |
Silver Care Plan, 1-yr Extended Warranty, Readouts |
SC2-THRO |
Silver Care Plan, 2-yr Extended Warranty, Readouts |
SC4-THRO |
Silver Care Plan, 4-yr Extended Warranty, Readouts |
Specifications |
1502A | 1504 |
Temperature Range† | −200 °C to 962 °C (−328 °F to 1764 °F) | Any thermistor range |
Resistance Range | 0 Ω to 400 Ω, auto-ranging | 0 Ω to 1 MΩ, auto-ranging |
Probe | Nominal RTPW: 10 Ω to 100 Ω RTD, PRT, or SPRT |
Thermistors |
Characterizations | ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 IPTS-68: R0, a, d, a4, and c4 Callendar-Van Dusen: R0, a, d, and b |
Steinhart-Hart thermistor polynomial Callendar-Van Dusen: R0, a, d, and b |
Resistance Accuracy (ppm of reading) |
0 Ω to 20 Ω: 0.0005 Ω 20 Ω to 400 Ω: 25 ppm |
0 Ω to 5 KΩ: 0.5 Ω 5 KΩ to 200 KΩ: 100 ppm 200 KΩ to 1 MΩ: 300 ppm |
Temperature Accuracy† | ±0.004 °C at −100 °C ±0.006 °C at 0 °C ±0.009 °C at 100 °C ±0.012 °C at 200 °C ±0.018 °C at 400 °C ±0.024 °C at 600 °C |
±0.002 °C at 0 °C ±0.002 °C at 25 °C ±0.004 °C at 50 °C ±0.010 °C at 75 °C ±0.020 °C at 100 °C (Using 10 KW thermistor sensor, a=0.04. Does not include probe uncertainty or characterization errors.) |
Operating Temperature Range | 16 °C to 30 °C | 13 °C to 33 °C |
Resistance Resolution | 0 Ω to 20 Ω: 0.0001 Ω 20 Ω to 400 Ω: 0.001 Ω |
0 Ω to 10 KΩ: 0.01 Ω 10 KΩ to 100 KΩ: 0.1 Ω 100 KΩ to 1 MΩ: 1 Ω |
Temperature Resolution | 0.001 °C | 0.0001 °C |
Excitation Current | 0.5 and 1 mA, user selectable, 2 Hz | 2 and 10 mA, automatically selected |
Measurement Period | 1 second | |
Digital Filter | Exponential, 0 to 60 seconds time constant (user selectable) | |
Probe Connection | 4-wire with shield, 5-pin DIN connector | |
Communications | RS-232 serial standard IEEE-488 (GPIB) optional |
|
Display | 8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters | |
Power | 115 VAC (±10 %), 50/60 Hz, 1 A, nominal 230 VAC (±10 %), 50/60 Hz, 1 A, nominal, specify |
|
Size (HxWxD) | 61 x 143 x 181mm (2.4 x 5.6 x 7.1 in) | |
Weight | 1.0 kg (2.2 lb.) | |
Probes from Fluke Calibration |
5615, 5627, 5626, 5628, 5622 | 5640-44, 5610-65 |
Calibration | Accredited NIST-traceable calibration provided | |
†Temperature ranges and accuracy may be limited by the sensor you use. |
Be the first to review “V Module”